Abstract

Since 2005 the Swiss Light Source (SLS) has been operating a Long Trace Profiler (LTP)-V from Ocean Optics in its metrology laboratory to measure the synchrotron optics for SLS. In 2012 we finished a significant upgrade to improve the accuracy, reliability and measurement efficiency in particular for the calibration of adaptive optics. Folding mirrors with figure errors <λ/100 and an additional linear encoder have been installed, the 1d CCD detector with 2048 pixels has been replaced by a 16 mega-pixel CCD camera with gigabit ethernet interface GigE, the monolithic software has been replaced by a modular, full- EPICS compatible system based on a new LTP plugin for the areaDetector software for image processing. The plugin allows slope determination in real time i.e. per frame.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.