Abstract

Data on several standard diffusion pumps indicate that the speed for hydrogen may be considerably less than 3.8 times the rated speed for air depending on the jet design, pump fluid, heater input, and speed of the fore pump. Evidence is presented that this reduction in speed is partly the result of diffusion of hydrogen back through the vapor jet from fore vacuum to high vacuum and partly the result of diffusion of hydrogen back out of the jet after entering from the high vacuum side. Measurements of the back diffusion of hydrogen, helium, and air in an experimental pump are correlated with the diffusion coefficient through an extension of the theories of Gaede and Matricon. The mechanism of jet breakdown has been observed with the aid of a glow discharge and a pilot dye.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.