Abstract
The pressed electrode techniques using graphite perwder as a conducting suppdit material have been investigated to obtain favorable source conditions for quantitative analysis of powder samples by spark source mass spectrographic method. 1. Using the pressing tool which we have developed, a rod sufficiently strong to be clamped in the electrode holders could be formed from mixtures of graphite and sample in any ratio, so far as we tested with KCl, CdS, ZnCl2and GaAs. 2. The relative sensitivity, for example, of K to Cl in the graphite-KCl pressed electrodes was found strongly dependent on the spark gap width. However, comparatively reproducible sensitivity(relative standard deviation less than10per cent)could be obtained, provided that the spark gap could be easily kept at about100μ, with rf spark voltage at 40-60 KV, repetition rate at 100-300 per second and the graphite content larger than ca50per cent by weight. 3. A correction method for impurity elements contained in graphite powder used as support material was proposed.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.