Abstract

Reproducibility of ion intensities, relative senisitivities of various elements by a spark source mass spectrograph were invesigated, and iron and steel were analyzed. Reproducibilities in repeated measurements with constant source parameters were5-35% in coefficients of variation, and in separate measurements6-70%. It seemed that inhomogeneity of emulsion and samples did not attribute to the fluctuation. Relative sensitivities(platinum was unity)of various elements in low alloy steel, nickel base temperature alloy, copper alloy, aluminum alloy, zinc base die casting alloy and stainless steel were 0.17-24.6. They were not related with sample materials, but with heat of sublimation of each element, however, some problems in reproducibility and measuring condition remain to be resolved. More accurate determination was performed by calibrating with relative sensitivities in analysis of standard samples stainless steel. Commercial pure iron was analyzed, and comparison with results in wet chemical methods was tried. The limit of detection was0.005 Reproducibility of ion intensities, relative senisitivities of various elements by a spark source mass spectrograph were invesigated, and iron and steel were analyzed.Reproducibilities in repeated meas urements with constant source parameters were5-35%in coefficients of variation, and in separate measurements6-70%. It seemed that inhomogeneity of emulsion and samples did not attribute to the fluctuation.Relative sensitivities(platinum was unity)of various elements in low alloy steel, nickel base temperature alloy, copper alloy, aluminum alloy, zinc base die casting alloy and stainless steel were0.17-24.6.They were not related with sample materials, but with heat of sublimation of each element, however, some problems in reproducibility and measuring condition remain to be resolved. More accurate determination was performed by calibrating with relative sensitivities in analysis of standard samples stainless steel. Commercial pure iron was analyzed, and comparison with results in wet chemical methods was tried. The limit of detection was0.005-1ppm.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.