I investigated the "skew ray" issue with respect to ocular aberration measurement. This arises for aberrometers that can distinguish only radial components of aberrations. Using simulations and Zernike aberrations, I found that rotationally symmetric coefficients are correctly determined but that asymmetric coefficients are underestimated. This underestimation is more pronounced as radial order of the aberrations becomes smaller and as angular dependence of aberrations becomes greater. There is also "leakage" to other orders, and errors are slightly affected by the sampling pattern. Aberrometers that can distinguish only radial components of aberrations do not provide good estimates of asymmetric aberrations.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.