Abstract

A rapid radioisotope induced energy-dispersive X-ray fluorescence procedure for the simultaneous determinations of Cobalt, Nickel and Molybdenum in hydrodesulfurization catalysts is presented. A109Cd (7 mCi) source is used as the excitation system coupled to a high resolution Si(Li) detector. A PDP-11/05 processor is employed as a multichannel analyzer and also to automatically control the complete system by means of a prewritten computer program. The relative standard deviations for the three determinations (CoO, NiO and MoO3) are less than 5% for 300 seconds of fluorescent time for a typical sample. Finally a comparison of results by different methods is given to check the accuracy since no standard reference materials are available for catalytic material.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.