Abstract
Abstract The separation of the three major isomers and the main photoproducts of bilirubin by Ion-Pair High Pressure Liquid Chromatography is described. Chromatographic procedures found in the literature proved difficult to reproduce, had unacceptably long elution times (over 2 hours), required complex gradients or two or more columns. As these reports provided no data to show why specific chromatographic conditions were chosen an optimization study was carried out. The separation is shown to be dependent on pH, organic content and buffer concentration of the mobile phase, ion-pairing agent concentration and injection solvent. A separation, carried out in less than thirty minutes on octadecyl silica using isocratic conditions with detection at 455 nm is demonstrated. This technique will be used in the elucidation of the mechanism involved in phototherapy treatment of neonatal jaundice.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.