Abstract

This study ascertained the viability of a portable liquid chromatograph, operating at low microliter per minute flow, for the analysis of seized drugs at remote sites as well as in laboratory settings. Synthetic cathinones were screened using dual capillary columns in series, C8 and biphenyl, with on-column ultraviolet detection at 255 and 275nm. The relative retention times of the two columns in series and their peak area absorbance ratio were used to determine if the 16 synthetic cathinones investigated could be uniquely identified in these conditions. Based on these parameters all of the analytes could be distinguished. Representative mixtures of synthetic cathinones were used to determine the repeatability, linearity, and limits of detection of the method. This cost effective and green instrumentation has the potential to satisfy minimum international guidelines for the analysis of seized drugs.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.