Abstract

Multiparameter optimizations have been carried out to study the effects of technological advances on the patient doses required to maintain a given image quality in mammography. The assumed advances include; improvement in the power loading limits of the tube focal spot, increased absorption efficiency for a given detector resolution, detector system gain increases and changes in the exposure time limitations that result from patient motion. The optimization permits system geometry, kVp of the examination, filtration, detector resolution, focal spot size and grid characteristics to vary simultaneously and self-consistently subject to image quality as well as technological constraints. In this study the technological constraints are systematically varied so that the effect of technological innovations can be measured by the resulting dose reductions relative to a baseline optimized system.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.