Abstract

Analytical expressions have been written for image quality in mammography. Multiparameter optimizations have been conducted to find mammography systems providing the lowest patient dose for a given image quality. The optimizations are subject to constraints imposed by technology, such as power limits on the tube focal spot, absorption efficiency related to detector resolution, and others. The optimizations permit system geometry, kVp, filtration, detector resolution, focal spot size, and grid characteristics to vary simultaneously and self-consistently subject to the constraints. A system configuration approaching a factor of 3 dose reduction has been found without assuming radical technological advances. The system satisfies image quality constraints for both large and small targets and would be possible to implement clinically. The sensitivity of the results to the assumptions made in the modeling has been investigated.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.