Abstract

Technological advances in HLA laboratory testing undoubtedly improved the sensitivity and specificity of HLA antibody assessment but not without introducing a set of challenges regarding data interpretation. In particular, the introduction of solid-phase single-antigen bead (SAB) antibody assessment brought the belief that mean fluorescence intensity (MFI) was a quantifiable value. As such, MFI levels heavily influenced HLA antibody reporting, monitoring, and clinical practice. However, given that SAB testing was neither intended for nor approved to be quantifiable, is the use of MFI in current clinical and laboratory practice valid? What, if anything, does this numerical value actually reveal about the pathogenic potential of the antibody? What are the pitfalls and caveats associated with reporting MFI? Herein, we travel the road to HLA antibody assessment and explore the reliability of MFI values to make clinical decisions.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.