Abstract

The paper presents a theoretical model for the response of a tissue-equivalent proportional counter (TEPC) irradiated with charged particles. Heavy ions and iron ions in particular constitute a significant part of radiation in space. TEPCs are used for all space shuttle and International Space Station (ISS) missions to estimate the dose and radiation quality (in terms of lineal energy) inside spacecraft. The response of the tissue-equivalent proportional counters shows distortions at the wall/cavity interface. In this paper, we present microdosimetric investigation using Monte Carlo track structure calculations to simulate the response of a TEPC to charged particles of various LET (1 MeV protons, 2.4 MeV alpha particles, 46 MeV/nucleon 20Ne, 55 MeV/nucleon 20Ne, 45 MeV/nucleon 40Ar, and 1.05 GeV/nucleon 56Fe). Data are presented for energy lost and energy absorbed in the counter cavity and wall. The model calculations are in good agreement with the results of Rademacher et al. (Radiat. Res. 149, 387-389, 1998), including the study of the interface between the wall and the sensitive region of the counter. It is shown that the anomalous response observed at large event sizes in the experiment is due to an enhanced entry of secondary electrons from the wall into the gas cavity.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call