Abstract

The major property of radiophotoluminescent glass dosimeter is its repeatable readout for the radiation measurements. Accordingly, the radiophotoluminescent glass dosimeter system will become one of the important dosimeters for dose measurement in the future. The aims of this study were to develop and characterize the self-fabricated radiophotoluminescent glass dosimeter reader system. The photomultiplier tube, charge-coupled device, nitrogen pulsed laser, band pass filter, 500 nm long pass filter and GD-302 glass dosimeter were used in this study. Based on our study, we found that the values of R-squared for both photomultiplier tube and charge-coupled device in reading system were 0.9968 and 0.9992, respectively, which shows that the relationship between luminescent signals or fluorescence images intensity of glass dosimeter and irradiation dose is direct proportion. These results suggest that the photomultiplier tube and charge-coupled device optical detector can be used in radiophotoluminescent glass dosimeter readout system.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.