Abstract

Effect of the pneumatic tube system (PTS) on sample quality is controversial. Herein we aim at evaluating the impact of sample transportation via the PTS on complete blood count (CBC) results. Duplicate CBC samples from normal donors and anemic patients were sent in parallel to the laboratory for testing through the PTS and the courier (CO). We used scatter plots, Bland-Altman plots, correlation coefficient (r), and coefficient of determination for the validation. A total of 115 samples (donors: 59, patients: 56) were tested. There was excellent correlation between both methods for red blood cell parameters (r range=0.9213-0.9958) and platelet count. White blood cell (WBC) count and differential count showed similar results (r range=0.8605-0.9821) for all, with exception of basophils which showed modest correlation (r=0.4827 for patients and 0.5758 for normal donors). Most of the differences in measurement of all CBC parameters were within the 95% confidence interval of the mean difference on Bland-Altman plots. Modern PTS can be safely used for transporting CBC samples.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.