Abstract
A two-dimensional parallel gap proportional counter has been developed at the Danish Space Research Institute. Imaging over the 120 mm diameter active area is obtained using the positive ion component of the avalanche signals as recorded by a system of wedge- and strip-electrodes. An electronically simple, but very effective background rejection is obtained by using the fast electron component of the avalanche signal. Gas gains up to 8 × 10 5 have been achieved. An energy-resolution of 16% and a sub-millimeter spatial resolution have been measured at 5.9 keV for an operating gas gain of 10 5. In principle, the position coordinates are linear functions of electronic readouts. The present model, however, exhibits non-linearities, caused by imperfections in the wedge and strip-electrode pattern. These non-linearities are corrected by using a bilinear correction algorithm. We conclude that the rugged construction, the simple electronics, the effectiveness of the background rejection and the actual imaging performance makes this a very attractive laboratory detector for low and intermediate count rate imaging applications.