Abstract

The effect of patient shielding on fetal radiation dose was evaluated in computed tomography pulmonary angiography with the out-of-plane shield visible in the localizer but absent in the scan range in chest computed tomography (CT). An anthropomorphic phantom with additional prosthetic pregnancy belly was scanned with different CT scanners using clinical imaging protocols and radiophotoluminescence dosemeters (GD-352M). The out-of-plane shield decreased the fetal absorbed radiation dose with Siemens Somatom go.Up, Canon Aquilion Prime SP and Canon Aquilion One scanners. The decrease was 3.9%-39.4% (0.01-0.09mGy). With GE Optima the shield increased the fetal dose by 100% (0.23mGy), with Canon Aquilion One and GE Optima scanners the abdomen dose increase was 17.5% and 36.4%, respectively (0.61 and 1.38mGy). Applying an out-of-plane shield outside the scanned volume may increase the fetal radiation dose during CT when using tube current modulation, depending on the make and model of the CT scanner.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.