Abstract

An optimization analysis of digital fluoroscopy has been carried out using (signal/noise)2/dose as a figure of merit. Three signal detection tasks have been considered: (i) small iodine-containing blood vessels; (ii) regions of fat; and (iii) microcalcifications. The effects of x-ray beam filtration and kilovoltage on the figure of merit have been parameterized and the associated x-ray tube power loadings have been defined. Results are presented in a fashion that facilitates the choice of parameters (amount of aluminum filtration, kilovoltage, and x-ray tube power loading) such that an acceptable balance can be achieved between signal detection sensitivity, patient dose, and x-ray tube power loading.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.