Abstract

Background Esophageal pressure is used as a reliable surrogate of the pleural pressure. It is conventionally measured by an esophageal balloon placed in the lower part of the esophagus. To validate the correct position of the balloon, a positive pressure occlusion test by compressing the thorax during an end-expiratory pause or a Baydur test obtained by occluding the airway during an inspiratory effort is used. An acceptable catheter position is defined when the ratio between the changes in esophageal and airway pressure (∆Pes/∆Paw) is close to unity. Sedation and paralysis could affect the accuracy of esophageal pressure measurements. The aim of this study was to evaluate, in mechanically ventilated patients, the effects of paralysis, two different esophageal balloon positions and two PEEP levels on the ∆Pes/∆Paw ratio measured by the positive pressure occlusion and the Baydur tests and on the end-expiratory esophageal pressure and respiratory mechanics (lung and chest wall).MethodsTwenty-one intubated and mechanically ventilated patients (mean age 64.8 ± 14.0 years, body mass index 24.2 ± 4.3 kg/m2, PaO2/FiO2 319.4 ± 117.3 mmHg) were enrolled. In step 1, patients were sedated and paralyzed during volume-controlled ventilation, and in step 2, they were only sedated during pressure support ventilation. In each step, two esophageal balloon positions (middle and low, between 25–30 cm and 40–45 cm from the mouth) and two levels of PEEP (0 and 10 cmH2O) were applied. The ∆Pes/∆Paw ratio and end-expiratory esophageal pressure were evaluated.ResultsThe ∆Pes/∆Paw ratio was slightly higher (+0.11) with positive occlusion test compared with Baydur’s test. The level of PEEP and the esophageal balloon position did not affect this ratio. The ∆Pes and ∆Paw were significantly related to a correlation coefficient of r = 0.984 during the Baydur test and r = 0.909 in the positive occlusion test. End-expiratory esophageal pressure was significantly higher in sedated and paralyzed patients compared with sedated patients (+2.47 cmH2O) and when esophageal balloon was positioned in the low position (+2.26 cmH2O). The esophageal balloon position slightly influenced the lung elastance, while the PEEP reduced the chest wall elastance without affecting the lung and total respiratory system elastance.ConclusionsParalysis and balloon position did not clinically affect the measurement of the ∆Pes/∆Paw ratio, while they significantly increased the end-expiratory esophageal pressure.

Highlights

  • Esophageal pressure is used as a reliable surrogate of the pleural pressure

  • Esophageal and airway pressure ratio During the positive pressure occlusion test, the pressure manually applied on the rib cage generated average positive changes in esophageal pressure of 6.9 ± 2.0 cmH2O and of 7.0 ± 2.0 cmH2O at 0 and 10 cmH2O of positive end-expiratory pressure (PEEP) respectively, while during the Baydur test the inspiratory muscle

  • Paralysis did not affect the changes in esophageal pressure at 0 and 10 cmH2O of PEEP (p = 0.325)

Read more

Summary

Introduction

Esophageal pressure is used as a reliable surrogate of the pleural pressure. It is conventionally measured by an esophageal balloon placed in the lower part of the esophagus. To validate the correct position of the balloon, a positive pressure occlusion test by compressing the thorax during an end-expiratory pause or a Baydur test obtained by occluding the airway during an inspiratory effort is used. The aim of this study was to evaluate, in mechanically ventilated patients, the effects of paralysis, two different esophageal balloon positions and two PEEP levels on the ∆Pes/∆Paw ratio measured by the positive pressure occlusion and the Baydur tests and on the end-expiratory esophageal pressure and respiratory mechanics (lung and chest wall). The conventional test to validate the correct position of esophageal balloon is performed during an end-expiratory occlusion maneuver with the simultaneous measurement of the changes in airway and esophageal pressure during an inspiratory effort (i.e., no air flow) (Baydur’s occlusion test) [12]. Milner et al [14] found in spontaneous breathing babies that esophageal pressure well correlated with airway pressure supporting the validity of this occlusion test in infants too

Objectives
Methods
Results
Discussion
Conclusion