Abstract

A new detector array, the Modular Total Absorption Spectrometer (MTAS), was commissioned at the Holifield Radioactive Ion Beam Facility (HRIBF) at Oak Ridge National Lab (ORNL). Total absorption gamma spectra measured with MTAS are expected to improve beta-feeding patterns and beta strength functions in fission products. MTAS is constructed out of hexagonal NaI(Tl) detectors with a unique central module surrounded by 18 identical crystals assembled in three rings. The total NaI(Tl) mass of MTAS is over 1000kg. The response of the central and other 18 MTAS modules to γ-radiation was simulated using the GEANT4 toolkit modified to analyze the nonlinear light output of NaI(Tl). A detailed description of the GEANT4 modifications made is discussed. Simulated energy resolution of MTAS modules is found to agree well with the measurements for single γ transitions of 662keV (137Cs) with 8.2% full width half maximum (FWHM), 835keV (54Mn) with FWHM of 7.5% FWHM, and 1115keV (65Zn) with FWHM of 6.5%. Simulations of single and multiple γ-rays from 60Co are also discussed.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.