Abstract
We have studied the response of a high resolution Ti/Au transition edge sensor (TES) microcalorimeter with an energy resolution of 6.2±0.7eV at 1.5keV and 7.8±0.9eV at 6.4keV. We find that the sensitivity α≡R/TdR/dT of the TES varies significantly along the superconducting-to-normal transition that the bias point traverses during a pulse. Furthermore, α is reduced significantly during a pulse compared to its equilibrium value calculated from the DC characteristics. This leads to an overestimate of the expected energy resolution when basing the prediction on the equilibrium sensitivity.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.