Abstract
The performance of X-ray microcalorimeters with a Ti/Au Transition Edge Sensor (TES) and a Cu absorber has been investigated. With these microcalorimeters we achieve energy resolutions below 4.5 eV FWHM for 5.89 keV X-rays in combination with a response time of about 100 μs. The measured energy resolution is analyzed and discussed based on spectral and RMS noise measurements. Excess noise was present which, in one case, could be fitted by noise proportional to 1α, suggesting the presence of thermal fluctuations in the TES. Variation in geometry and illumination was examined but did not influence the energy resolution.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Similar Papers
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.