Abstract

We describe how the National Institute of Standards and Technology obtains a scale of absolute spectral response from 406 nm to 920 nm. This scale of absolute spectral response is based solely on detector measurements traceable to the NIST High Accuracy Cryogenic Radiometer (HACR). Silicon photodiode light-trapping detectors are used to transfer optical power measurements from the HACR to a monochromator-based facility where routine measurements are performed. The transfer also involves modeling the quantum efficiency (QE) of the silicon photodiode light-trapping detectors. We describe our planned quality system for these measurements that follows ANSI/NCSL Z540-1-1994. A summary of current NIST capabilities based on these measurements is also given.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.