Abstract

In 24-hour Non-invasive Blood Pressure (NIBP) measurement, most of the automatic devices use oscillometric method. Accurate measurement of BP at finger using an Oscillometric Waveform (OMW) remains a challenge, particularly when the waveform is collected in inflation, because of the bad signal-to-noise ratio. In the present work, we aimed at measuring Ambulatory Blood Pressure by extracting OMW signal at finger with a new method of linear inflation control and digital signal processing. With the PID control of the pump and real time signal processing, we implemented the blood pressure measurement system and new algorithm. For the validation of our suggested method, we compare the effect of our method and traditional method. Finally, the method is validated in a pilot study against Welch Allyn ProBP 3400 BP monitors.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.