In the X-ray single-grating imaging system, the acquisition of frequency information is the key step of phase-contrast and scattering information recovery. In the process of information extraction, it is easy to lead to the degradation of imaging quality due to the Moire Artifact, thus limiting the development and application of X-ray single-grating imaging system. In order to address the above problems, in this article, based on the theoretical analysis of the generation principle of Moire Artifact in imaging system, the advantages and disadvantages of grating rotation method are analyzed, and a method of suppressing Moire artifacts by adjusting grating projection frequency is proposed. The experimental results show that the method proposed here can suppress the Moire noise in the background noise, resulting in a reduction of more than 50% in the standard deviation of the background noise. High quality phase-contrast and scattering images are obtained experimentally, which is of great value to the development of X-ray single-grating imaging technology.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.