An alpha/beta emission rate counting system which comprises of a large-area windowless multi-wire proportional counter and a digital pulse processing module has been established at the National Institute of Metrology of China, aiming at realizing the absolute measurement of emission rate for alpha and beta sources. Performance of the detector, including high-voltage plateaus and uniformity were measured. Measured emission rates for large-area 241Am and 90Sr/90Y sources when compared with certificated values showed difference within 1.0%, and good agreements achieved.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.