Abstract

A large area windowless gas flow multi wire proportional counting system for the calibration of large area reference sources has been developed as a primary standard at Bhabha Atomic Research Centre (BARC). The counting system consists of a multi wire proportional counter (MWPC), vacuum system, gas flow system and pulse processing units. The MWPC detector assembly consists of a vacuum tight aluminum enclosure, multi wire grid and sliding source tray. Various detector characteristics like operating characteristics curve, Fe-55 spectrum for beta discriminator threshold setting and dead time of the measurement system were studied and determined in order to achieve an optimized detection capability. The surface emission rates of different source strengths were measured and their relative combined standard uncertainties were determined. Large Area Sources Comparison Exercise (LASCE) was organized by International Committee on Radionuclide Metrology (ICRM) working group and coordinated by National Institute for Ionising Radiation Metrology (ENEA), Italy, to demonstrate equivalence of surface emission rate measurements at the international platform. BARC participated in the programme and the results of LASCE are also discussed in this paper.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.