Abstract

The ED-XRF (Energy-Dispersive X-ray Fluorescence) measurements and the FLUKA simulations have been made to discuss the possibility of recognition of coating layer as well as to its thickness measurement. In this work the IKα(Cu)/IKα(Ag) intensity ratios as well as IKβ(Ag)/IKα(Ag) and IKβ(Cu)/IKα(Cu) for copper samples coated with various thickness of sputtered silver have been analyzed. The results show strong dependence of these factors with coating silver layer thickness.The measurements show the performance of this method in archaeometry. Since the use of non-destructive methods during tests on ancient silver artifacts may not supply to obtain reliable bulk results and should be considered applicable for only surface analyses, the measured intensity ratios can be applied as a tool to estimation of surface silver enrichment thickness.