Abstract

To be accurate and equivalent, laboratory results should be traceable to higher-order references. Furthermore, their quality should fulfill acceptable measurement uncertainty (MU) as defined to fit the intended clinical use. With this aim, in vitro diagnostics (IVD) manufacturers should define a calibration hierarchy to assign traceable values to their system calibrators. Medical laboratories should know and verify how manufacturers have implemented the traceability of their calibrators and estimate the corresponding MU on clinical samples. Accordingly, the internal quality control (IQC) program should be redesigned to permit IVD traceability surveillance through the verification by medical laboratories that control materials, provided by the manufacturer as a part of measuring systems, are in the clinically suitable validation range (IQC component I). Separately, laboratories should also monitor the reliability of employed IVD measuring systems through the IQC component II, devoted to estimation of MU due to random effects and to obtaining MU of provided results, in order to apply prompt corrective actions if the performance is worsening when compared to appropriate analytical specifications, thus jeopardizing the clinical validity of test results.