Abstract

This study aimed to evaluate discrepancies in potassium measurements between point-of-care testing (POCT) and central laboratory (CL) methods, focusing on the impact of hemolysis on these measurements and its impact in the clinical practice in the emergency department (ED). A retrospective analysis was conducted using data from three European university hospitals: Technische Universitat Munchen (Germany), Hospital Universitario La Paz (Spain), and Erasmus University Medical Center (The Netherlands). The study compared POCT potassium measurements in EDs with CL measurements. Data normalization was performed in categories for potassium levels (kalemia) and hemolysis. The severity of discrepancies between POCT and CL potassium measurements was assessed using the reference change value (RCV). The study identified significant discrepancies in potassium between POCT and CL methods. In comparing POCT normo- and mild hypokalemia against CL results, differences of-4.20 % and +4.88 % were noted respectively. The largest variance in the CL was a +4.14 % difference in the mild hyperkalemia category. Additionally, the RCV was calculated to quantify the severity of discrepancies between paired potassium measurements from POCT and CL methods. The overall hemolysis characteristics, as defined by the hemolysis gradient, showed considerable variation between the testing sites, significantly affecting the reliability of potassium measurements in POCT. The study highlighted the challenges in achieving consistent potassium measurement results between POCT and CL methods, particularly in the presence of hemolysis. It emphasised the need for integrated hemolysis detection systems in future blood gas analysis devices to minimise discrepancies and ensure accurate POCT results.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call