Abstract
Purpose. Comparison of computer tomography (СT) contrast enhancement (CE) protocols for optimal visualization of cardiac chamber, definition it’s influence on results of non-invasive superficial cardiac mapping.Materials and methods. The study included 93 patients with heart rhythm disorders who planned catheter ablation of arrhythmia. Noninvasive cardiac mapping was made for topical diagnostics. It includes multichannel ECG-registration and CT with intravenous СE (1st group monophasic (50 patients), 2nd group split-bolus (18 patients), 3rd group with pre-bolus (25 patients). Qualitative and quantitative (measurement of mean blood attenuation in four chambers, calculation of ventricular-myocardial contrast-to-noise ratio VM-LV и VM-RV for left ventricle (LV) and right ventricle (RV), respectively) parameters were compared between groups. Fusion of ECG and CT data was made semi-automatic with diagnostic complex «Amycard 01К».Results. Regardless of CE technique was noted sufficient and homogeneous contrast attenuation of left atrium (LA) and LV (mean blood attenuation in LA more than 278 HU, LV 250 HU, VM-LV 0,582). Enhancement of right heart was insufficient with monophasic protocol, in most cases the average CT density was lower than 200 HU, VM-RV 0,256. Split-bolus protocol improves visualization of right atrium (RA) and RV (blood density in RA 258HU, RV 227HU, VM-RV 0,541), however there was heterogeneity of RA cavity because of artifacts from superior vena cava (VC) and unenhanced blood from inferior inferior VC. Using of pre-bolus increases contrast ratio between RA myocardium and blood due to increasing CT density of blood in inferior VC (blood density 294 HU). Quality of right ventricle CE was similar to 2nd group (blood density 264 HU, VM-RV 0,565).Conclusion. CE protocols split-bolus and with pre-bolus improve visualization of right ventricle, supporting the high level enhancement of left heart. Protocol with pre-bolus is preferable for exact differentiation of right atrial endocardial contour.
Highlights
Контрастное усиление по методике сплит-болюс за счет удлинения времени введения позволило улучшить визуализацию правых отделов (плотность в правом предсердии (ПП) 258 HU, правого желудочка (ПЖ) 227 HU, VM-RV 0,541), однако сохранялась гетерогенность полости правого предсердия (ПП) за счет артефактов от верхней половой вены (ПВ) и притока неконтрастированной крови из нижней ПВ
Qualitative and quantitative (measurement of mean blood attenuation in four chambers, calculation of ventricular-myocardial contrast-to-noise ratio VM-LV и VM-RV for left ventricle (LV) and right ventricle (RV), respectively) parameters were compared between groups
Enhancement of right heart was insufficient with monophasic protocol, in most cases the average CT density was lower than 200 HU, VM-RV 0,256
Summary
The study included 93 patients with heart rhythm disorders who planned catheter ablation of arrhythmia. Noninvasive cardiac mapping was made for topical diagnostics. It includes multichannel ECG-registration and CT with intravenous СE (1st group - monophasic (50 patients), 2nd group - split-bolus (18 patients), 3rd group with pre-bolus (25 patients). Qualitative and quantitative (measurement of mean blood attenuation in four chambers, calculation of ventricular-myocardial contrast-to-noise ratio VM-LV и VM-RV for left ventricle (LV) and right ventricle (RV), respectively) parameters were compared between groups. Fusion of ECG and CT data was made semi-automatic with diagnostic complex «Amycard 01К»
Published Version (Free)
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have