Abstract

We have determined limits of detection (LODs) of DNA hybridization using an optical detection by two different method, i.e., surface plasmon fluorescence spectroscopy (SPFS) and optical waveguide fluorescence spectroscopy (OWFS), however, an identical sensor. Both techniques have a similar detection principle using the enhanced electromagnetic field to excite fluorophores and to monitor DNA hybridization in real-time with high sensitivity. Firstly, the detection of DNA hybridization was conducted by SPFS and the LOD value was found to be 2.0pM (2.0×10−12mol∕L). The same study was further conducted by OWFS, and a LOD of 170fM (1.7×10−13mol∕L) and 100fM (1×10−13mol∕L) was achieved for TM1 and TE0 modes, respectively, indicating that an improvement of the LOD of one order of magnitude is possible with OWFS. Furthermore, the photonic mode density at resonance of each mode was calculated, and these values are reflected in the LOD values. The present study illustrates the potential of OWFS as biosensor applications being able to detect in real-time even trace amounts of analyte.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.