Abstract
A front-end electronics of dose monitor has been developed for measuring irradiation dose to the patient in Shanghai Advanced Proton Therapy Facility. The parallel plate ionization chamber is used for the dose monitoring. Unlike the traditional method of recycling capacitor integration and voltage-to-frequency conversion, this dose monitor electronics uses the trans-impedance amplifier and analog-to-digital conversion method. It performs satisfactorily, with the integral nonlinearity of less than ±0.04 nA in the range of −400 to 50 nA and the resolution of about ±0.6 nA.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.