Abstract
A chemiluminescence (CL) system that combines the simplicity and reproducibility of sequential injection analysis (SIA) with the radial flow properties of the fountain cell was developed. Optimum conditions for SIA involving three overlapping zones were deduced and the analytical utility of the system was demonstrated by the determination of hydrogen peroxide and glucose, both analyses based on the luminol reaction. The results obtained show that the combination of SIA and the fountain cell can be used for quantitative analysis and a study of reaction kinetics. The system has minimum reagent consumption and can be evaluated without any reconfiguration of the equipment or reagents involved in a given analysis.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.