Abstract

The presence of an intra-articular pump has been proposed as a central mechanism in the process of osteolysis and aseptic loosening of hip arthroplasty. It is not known if this pump exists and its mechanism remains uncharacterised. This study describes a new in vitro model of a cemented femoral stem in which cement/stem interface fluid pressures can be reliably measured under dynamic loads simulating stair climbing. A stem pump mechanism was found that generates both positive and negative clinically significant pressures (mean pressure ranges 5000–17,000 Pa). The timing of pressure peaks on the anterior and posterior aspects of the stem were in anti-phase, giving rise to oscillatory pressure gradients and potentially generating oscillatory fluid flows during the simulated physiological load cycle. The pump mechanism was shown to occur at the interface of a newly implanted polished double-tapered stem and emphasizes the importance of a complete mantle to protect the femoral bone from the raised fluid pressures.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.