Abstract

Background: To evaluate contrast-to-noise ratio (CNR) measurements in assessing image quality, in the context of the detectability performance of low-contrast detail (LCD), in computed tomography (CT) images, since exposure to elevated ionising-type radiation is considered to present excessive carcinogenic risk, whilst also causing distress in study subjects. Methods: An LCD phantom module (CTP515) was utilised in the study. Three dissimilar contrast items were used to analyse the ramifications of the proportions of an object on the CNR. Three multidetector CT (MDCT) scanners were used, with 16-MDCT, 64-MDCT and 80-MDCT frameworks, respectively. The CT scans were recreated using three dissimilar remaking algorithms—soft, standard and lung. The effects exerted on the CNR by various remodelling algorithms, as well as the contrast of various objects along with the size of the objects, were explored. The Hounsfield units of each chosen object (one unit representing the outer portion of the object) and the background and the standard deviation of the noise parameter were quantified, and algorithms were developed using MATLAB. Results: The CNR information was greatly influenced by changing the image recreation calculations and was very much increased in the soft-tissue recreation images using 16-MDCT and 64-MDCT. The CNR information was also increased more in the optimum recreation images than in the reproduced images from the computational procedure used in the 80-MDCT. The results did not show any remarkable contrasts in the CNR values between the different object sizes. Overall, a higher kVp produced an improved CNR in all the CT scanners. In particular, there were prominent upgrades in the CNR information when the kVp was increased from 80 to 120. Higher mAs levels gave better CNR values overall, especially for greater section thicknesses. Based on the CNR estimations, the 64-MDCT provided the best correlation among the CT scanners. Conclusions: The objective LCD appraisal method, based on CNR measurements, was confirmed as being useful for checking the different impacts of kVp, mAs and section thickness on the nature of the picture. This procedure was similarly viable in assessing the impacts of the different reconstruction calculations and the different differentiation questions on the nature of the image.

Highlights

  • A significant increase in the number of computed tomography (CT) scanners has highlighted the need to reduce the amount of radiation used in them [1]

  • We examined the differences between different CT scanners, in terms of the contrast-to-noise ratio (CNR) estimations of those scanned objects

  • The effects of the size of the objects, with 1% contrast level, on CNR values were evaluated for different multidetector CT (MDCT) scanners, reconstruction algorithms, object contrast levels and mAs selections (Figure 2 and Tables 3 and 4)

Read more

Summary

Introduction

A significant increase in the number of computed tomography (CT) scanners has highlighted the need to reduce the amount of radiation used in them [1]. One solution for improving image quality at low-radiation levels has relied on progress being made in the rebuilding methods [5,6,7,8], among which iterative reconstruction (IR). To evaluate contrast-to-noise ratio (CNR) measurements in assessing image quality, in the context of the detectability performance of low-contrast detail (LCD), in computed tomography (CT) images, since exposure to elevated ionising-type radiation is considered to present excessive carcinogenic risk, whilst causing distress in study subjects. The CNR information was increased more in the optimum recreation images than in the reproduced images from the computational procedure used in the 80-MDCT.

Objectives
Results
Discussion
Conclusion
Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.