Abstract

The f-ratio method is a standard-based quantification method which was originally developed based on a cold field emission scanning electron microscope/energy dispersive spectroscopy (CFE-SEM/EDS) system. It incorporates traditional EDS experiments and Monte Carlo simulations, and standards with known compositions are needed to calibrate the differences between them. However, as a new quantification method, the f-ratio method still requires extensive validations to be performed in different systems. In this study, the f-ratio method was used to analyze three certified minerals: kyanite (Al2SiO5), albite (NaAlSi3O8) and orthoclase (KAlSi3O8), and the option of standards was extended to any standard specimen containing the target elements. The effects of standards and beam energies on the calibration factors and the quantification results were discussed, and no significant influence was observed. The quantification results were also compared with the conventional standardless analysis and the standard-based k-ratio method, and the results show the f-ratio method is capable to achieve both the merits of a satisfactory accuracy and simple experimental procedures.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.