Abstract

Heart failure, a common clinical syndrome caused by functional and structural abnormalities of the heart, affects 64 million people worldwide. Long-term mechanical circulatory support can offer lifesaving treatment for end-stage systolic heart failure patients. However, this treatment is not without complications. This review covers the major complications associated with implantable mechanical circulatory support devices, including strokes, pump thrombosis and gastrointestinal bleeding. These complications were assessed in patients implanted with the following devices: Novacor, HeartMate XVE, CardioWest, Jarvik 2000, HeartMate II, EVAHEART, Incor, VentrAssist, HVAD and HeartMate 3. Complication rates vary among devices and remain despite the introduction of more advanced technology, highlighting the importance of device design and flow patterns. Beyond clinical implications, the cost of complications was explored, highlighting the difference in costs and the need for equitable healthcare, especially with the expected rise in the use of mechanical circulatory support. Future directions include continued improvement through advancements in design and technology to reduce blood stagnation and mitigate high levels of shear stress. Ultimately, these alterations can reduce complications and enhance cost-effectiveness, enhancing both the survival and quality of life for patients receiving mechanical circulatory support.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call