Abstract

Pulse arrival time (PAT), evaluated from electro-cardiogram (ECG) and photoplethysmogram (PPG) signals, has been widely used for cuff-less blood pressure (BP) estimation due to its high correlation with BP. However, the question of whether filtering the PPG signal impacts the extracted PAT values and consequently, the correlation between PAT and BP, has not been investigated before. In this paper, using data from 18 subjects, changes in the PAT values, and in the subject-specific PAT-systolic BP (SBP) correlation caused by filtering the PPG signal with variable cutoff frequencies in the range of 2 to 15 Hz are studied. For PAT extraction, three PPG characteristic points (foot, maximum slope and systolic peak) are considered. Results show that differences in the cutoff frequency can shift the PAT values and introduce a worst-case error of over 8.2 mmHg for SBP estimation, indicating that PPG signal filter settings can impact PAT-based BP estimations. Our study suggests that extracting the PAT from the maximum slope point of PPG signal filtered at 10 Hz provides the most stable correlation with SBP.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.