Abstract

Computer simulations have been carried out to study the effects of the experimental parameters when the mirage method has been applied to thermal diffusivity measurements of oriented polymer films. The parameters under study are the thermal diffusivity of the fluid surrounding the sample, the modulation frequency and the radius of the heating beam, the height and the radius of the probe beam, and the sample thickness and thermal diffusivity. Proposals for the optimum parameter values to maximize the measurement sensitivity for the sample diffusivity are made and the difficulties arising from the low diffusivity of the samples are described. It is also concluded that because the thermal properties of the fluid surrounding the sample have a strong contribution to the mirage signals, the signals do not include any simple feature corresponding to the sample diffusivity. Therefore it should be determined from the entire measurement data using regression methods.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.