Abstract

A comparative study of centrifugation and conductance methods for the estimation of cell volume fraction (ϕ) was performed to examine whether the strong forces exerted upon erythrocytes during centrifugation affect their volume, and the results are discussed in terms of erythrocyte deformability. Rabbit erythrocytes of four shapes (spherocytes, echinocytes, stomatocyte-like enlarged erythrocytes and discocytes) were prepared by controlling the pH of the suspending media. The packed cell volumes of the suspensions were measured by standard hematocrit determination methods using centrifugation in capillary tubes. Simultaneously, the same suspensions and their supernatants were used in dielectric spectroscopy measurements, and the low-frequency limits of their conductivities were used for the numerical estimation of ϕ. The hematocrit values of spherocytes and echinocytes were markedly less than the volume fractions obtained by the conductance method. Namely, the centrifugation reduced the cell volume. For enlarged erythrocytes and discocytes, however, the reduction of cell volume was not observed. These findings showed that ϕ obtained by the centrifugation method can be greatly affected by the deformability of the cells, but the level of the effect depends on the cell types. Consequently, ϕ obtained by the centrifugation method should be carefully interpreted.