Abstract

The oscillation of a cantilever in ultrahigh vacuum dynamic force microscopy is sometimes measured with the help of a Fabry–Pérot interferometer. We show that the photoinduced forces present in such an interferometer can artificially increase or decrease the effective quality factor of the cantilever. We examine this effect on the basis of a PtIr-coated rectangular silicon cantilever at temperatures between 7.7 and 30 K. By measuring resonance curves we demonstrate that the magnitude of the effective quality factor changes with cantilever-fiber distance, laser power, and temperature.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.