Abstract

Falls are a major cause of death and morbidity in older adults. In recent years many researchers have examined the role of wearable inertial sensors (accelerometers and/or gyroscopes) to automatically detect falls. The primary goal of such fall monitors is to alert care providers of the fall event, who can then commence earlier treatment. Although such fall detection systems may reduce time until the arrival of medical assistance, they cannot help to prevent or reduce the severity of traumatic injury caused by the fall. In the current study, we extend the application of wearable inertial sensors beyond post-impact fall detection, by developing and evaluating the accuracy of a sensor system for detecting falls prior to the fall impact. We used support vector machine (SVM) analysis to classify 7 fall and 8 non-fall events. In particular, we focused on the effect of data window size and lead time on the accuracy of our pre-impact fall detection system using signals from a single waist sensor. We found that our system was able to detect fall events at between 0.0625-0.1875 s prior to the impact with at least 95% sensitivity and at least 90% specificity for window sizes between 0.125-1 s.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.