Abstract

The objective of this study was to investigate the effects of weight-bearing (WB) condition on the kinematics of total knee arthroplasty. We investigated 17 patients (20 knees) implanted with a high-flexion posterior-stabilized total knee arthroplasty using 2- to 3-dimensional registration techniques. In vivo kinematics of dynamic deep knee flexion under WB and non-WB (NWB) conditions were compared. Six degree-of-freedom kinematics and articular contacts including post-cam contact were evaluated. At midflexion, femorotibial contact points were located significantly more anteriorly under NWB than WB conditions. As a result, post-cam engagement occurred significantly earlier under NWB than WB conditions. With NWB conditions, early engagement of the femoral cam engages at the top part of tibial post, which puts it at risk of jumping the tibial post particularly if there is instability in midflexion.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.