Abstract

During clinical usage, scaler tips may become worn and reduced in length. It is unknown what effect wear has on the magnitude of scaler tip vibrations when they are utilized under typical clinical loads. The aim of this investigation was to assess the effect of simulated wear on ultrasonic scaler tip displacement amplitude, using a scanning laser vibrometer. A Cavitron SPS (Dentsply) ultrasonic generator and three scaler insert designs (FSI-100, FSI-1000 and FSI-SLI-10S) were selected for the investigation. Tip vibration displacement amplitude was assessed unloaded and then contacting against tooth surfaces with loads of 0.5 and 1.0 N. Tips were then ground down by 1 mm and then 2 mm and scans were repeated. For all tips, load and length were found to be significant variables (p<0.0001). The scaler tips showed a fall in displacement amplitude with a reduction in tip length. However, all scaler tips showed variability in the amount oscillation that occurred. This was most pronounced with FSI-SLI-10S. This investigation demonstrated that tip wear could affect the performance of dental ultrasonic scaler inserts by reducing their vibration displacement amplitude. Clinicians should be aware of this variability, which may be significant enough to affect clinical procedures.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.