Abstract

A continue-wave cavity ring down experimental system is established surrounding the micro-sphere cavity. In the experiments of measuring micro-spheres cavity quality factor (Q) upon the system, it is found that photoelectric detectors with different response characteristic have different response to the optical signal which is shut off rapidly. The vibration noise caused by changes of external environment, meanwhile, can be transformed following the change of photoelectric detector properties. With the build of photoelectric detector response model, the mean-square deviation of repeated experiments under the same conditions is used for charactering level of the error caused by vibration noise. The influence of vibration noise to photoelectric detector is discussed in detail, and further analyzes the effect for the cavity ring down time extraction. Then the optimal photoelectric detector (Newfocus1434, 15V/W) is selected to optimize the experimental system. The level of error caused by vibration noise has been decreased from 0.191ns to 1.562ps. Finally, the Q value of micro-sphere cavity is calculated according to fitted value of cavity ring down time, and the experimental correctness is verified by the line-width law.