Abstract

The purpose of the present study was to investigate if the exposure could be reduced from the clinical setting (resulting in an effective dose of 8 microSv for a neonate of weight 0.7 kg and height 25 cm at a tube voltage of 90 kV) without negatively influencing the image quality for a dual-side readout technique computed radiography (CR) system in chest radiography of premature neonates. Chest radiographs of premature neonates were acquired with the double-side readout technique CR system. The images underwent simulated dose reduction in steps of 20 % to represent five different radiation dose levels. Four image quality criteria, related to the visibility of important anatomical structures, were used in a visual grading study where five experienced radiologists rated how well the criteria were fulfilled for all images. When reducing the radiation dose, a decrease in image quality could be observed already at the 80 % dose level for all the structures. The results indicate that a decrease in exposure from the clinically used setting affects the image quality negatively for the CR system.