Abstract

Serum (1→3) beta-D-glucan (BG) measurement is a useful test for systemic mycoses, and often used. On the other hand, various factors, including administration of intravenous immunoglobulins (IVIG) may cause false-positives. In the present study, we measured BG concentration of seven IVIG preparations with three lots respectively. BG levels varied with individual IVIG preparations (<3.0 - >300 pg/mL), and contamination from manufacturing processes was suspected. With serum BG concentration of clinical specimens obtained in Niigata University Medical & Dental Hospital, the difference between before and after administration of IVIG were calculated. The false-positive rate of BG due to IVIG administration was 9.8 %, and the positive predective value was reduced to 37.5%. Above all, administration of IVIG can complicate the BG test's interpretation, and caution is required.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.