Abstract
Comprehension of the biomechanical behaviour of orthopaedic implants is essential. This paper describes the development of an in vitro model to investigate the behaviour of femoral implants in the revision setting. The development of a femoral model and a bone graft substitute is described. The properties of human, bovine, ovine morselized bone graft, and a graft substitute were compared. On measuring hoop strain after impaction bone grafting there was no significant difference between the ovine bone graft and graft substitute with the size 1 Exeter stem. The results suggest that this bone graft substitute is a viable alternative for in vitro testing. The authors recommend the use of the graft substitute and the femoral model to predict femoral stem biomechanics.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.