Abstract

Impaction bone grafting using morcellised allograft can successfully restore bone stock in revision surgery. However, concerns exist regarding supply of bone and transmission of infection. Bone-graft extenders, such as tricalcium phosphate (TCP) and hydroxyapatite (HA), are used to minimise the use of donor bone. However, concerns exist around a reported increased risk of femoral fracture during impaction bone grafting with a 1:1 mixture of TCP/HA and morcellised bone graft (MBG) during impaction grafting in human cadaveric femora. Using a sawbones model, it was evaluated whether there was increased femoral cortical strain with a HA:MBG mixture during impaction grafting compared to MBG impacted at the same and a greater force. Subsequently the subsidence behaviour of the different graft mixes was compared by using a loaded femoral stem in an endurance test. It was demonstrated that the femora with the MBG:HA graft had greater cortical hoop stresses but improved subsidence behaviour compared to a graft composed of pure MBG impacted at the same force.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.