Abstract

‘Radiation-independent’ aiming systems for the placement of interlocking screws in intramedullary tibial nails have failed because of insertional related nail deformation. Prior to the development of these distal aiming systems, the variable position of the nails' interlocking holes after insertion must be known. In this study, the amount and direction of implant deformation of solid stainless-steel tibial nails (diameters of 8 and 9 mm, n = 20) were analyzed. Measurement of implant deflection (three translations, three angles) in the center of the distal transverse locking hole was performed with a three-dimensional-magnetic motion tracker system after nail insertion in paired human cadaver tibiae. The results showed combined mean lateral translation of the 8 and 9 mm nails ( n = 20) of −4.5 ± 3.5 mm (range: 14.3 mm) and dorsal translation −7.8 ± 5.8 mm (range: 19.2 mm). The combined mean rotational deformation about the longitudinal axis of the nail was 0.3 ± 0.7 (range: 2.4). The authors conclude that a simple aiming arm, mounted on the proximal nail end alone, is not sufficient to provide accurate distal aiming.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.